Light meets molecular motion
Molecules can absorb infrared energy at frequencies associated with their vibrations. In a transmission measurement, some radiation passes through the sample while some is absorbed. The changing balance across frequencies carries information about the material.
A Fourier transform infrared instrument, usually called FTIR, gathers an interferogram: a combined signal containing information about many frequencies. A mathematical transformation turns that signal into a spectrum. The instrument therefore connects a physical interaction with a representation that can be inspected.
Read the axes first
Infrared spectra commonly use wavenumber, expressed in reciprocal centimeters, to locate features. The other axis may show absorbance or transmittance. An absorption feature rises on an absorbance plot but falls on a transmittance plot, so the direction of a feature needs context.
A background measurement helps distinguish the sample's response from the measurement path. Comparing a sample spectrum with reference spectra can support identification, but the pattern as a whole matters. Several features may overlap, particularly in mixtures.
A clue needs a method
Sample thickness, preparation and the chosen sampling technique can affect a spectrum. A library match is useful evidence to examine, rather than a substitute for checking whether the reference and sample are comparable.
Estimating concentration requires an appropriate quantitative method and calibration. A taller feature does not, by itself, establish how much of a substance is present. Keep the question, sampling approach and comparison method together when reading a result.